Found 90 products for "

insertion tube

"
Jakość SonoScape EG-500 Elektroniczna Gastroskopia do naprawy fabryka

SonoScape EG-500 Elektroniczna Gastroskopia do naprawy

SonoScape EG-500 Electronic Gastroscopy For Repair ► Brief Description Inspection Results: 1. No image display, CMOS water ingress fault; 2. 20% of the light bundle is broken; 3. There are bite marks on the insertion tube in multiple places; 4. The biopsy tube is worn out and not smooth; 5. The ...

Jakość OLYMPUS TJF-Q180V Elektroniczny dwunastokąt do naprawy fabryka

OLYMPUS TJF-Q180V Elektroniczny dwunastokąt do naprawy

OLYMPUS TJF-Q180V Electronic Duodenoscope For Repair ► Brief Description Inspection Results: 1. Image display is available; 2. The light bundle brightness is acceptable; 3. The insertion tube is worn; 4. The light guide tube is worn; 5. The angle is insufficient, the snake bone, steel wire and helix ...

Jakość Olympus LTF-160 do naprawy fabryka

Olympus LTF-160 do naprawy

Olympus LTF-160 thoracoscope for repair ► Brief Description Fault: 1.with image, but big risk for CCD after leakage. 2. Light bundle broken completely. 3. Insert tube with little scratch. 4. Biopsy tube abrasion and leakage. 5. Wires and helix tube abrasion. 6. Grip broken, leakage. 7. All the ...

Jakość R Wolf 7326071 ureteroskop do naprawy fabryka

R Wolf 7326071 ureteroskop do naprawy

R Wolf 7326071 ureteroscope for repair ► Brief Description Fault: The image has spots,image bundle aging,light bundle broken;Insert tube aging;Biopsy channel worn and leakage;Bending section,wire and helix tube were worn. Solution: Replace the light bundle,image bundle,rubber,biopsy channel,helix ...

Jakość Olympus CYF V2 Cystoskop do naprawy fabryka

Olympus CYF V2 Cystoskop do naprawy

Olympus CYF V2 Cystoscope For Repair ► Brief Description Fault: Leakage. Solution: Replace the insert tube, light bundle, biopsy channel, bending section, wire, steel mesh, rubber. ►Problem & Solutoin Brand Olympus Model CYF V2 Fault Leakage Solution Replace the insert tube, light bundle, biopsy ...

Jakość Olympus A70940A Artroskop do naprawy fabryka

Olympus A70940A Artroskop do naprawy

Olympus A70940A Autoclavable Arthroscope for repair ► Brief Description Fault: 1. Image blur, black dots and dark. 2. Fiber not light enough. 3. Rod lens broken. 4. Objective lens broken and dirty. Solution: 1. Replace rod lens. 2. Replace inner tube. 3. Replace outer tube. 4. Replace objective lens ...

Jakość GIF-XQ240 Gastroskop od Olympus do naprawy fabryka

GIF-XQ240 Gastroskop od Olympus do naprawy

GIF-XQ240 Gastroscope by Olympus for repair ► Brief Description Fault: Leakage. Solution: Replace light guide tube,NO.1 button,wire,mesh,adjust angle. ►Problem & Solutoin Brand Olympus Model GIF-XQ240 Fault Leakage. Solution Replace light guide tube,NO.1 button,wire,mesh,adjust angle. ► Common ...

Jakość Kurs certyfikacji naprawy sprzętu medycznego Olympus 290 Kolonoskopia i bronchoskopia dla południowokoreańskiego klienta fabryka

Kurs certyfikacji naprawy sprzętu medycznego Olympus 290 Kolonoskopia i bronchoskopia dla południowokoreańskiego klienta

The courses we organize and their contents are as follows. COURSE: Olympus 290 Colonoscopy and Bronchoscopy Repair Training Course Brands: OLYMPUS, FUJINON Duration: 5 Days / 4 hours per day (20 hours total) Course Schedule: Courses run on demand. We open the course when you ask for it and arrange ...

Jakość Zespół cewki o zmiennej sztywności do kolonoskopu Olympus CF-H190L fabryka

Zespół cewki o zmiennej sztywności do kolonoskopu Olympus CF-H190L

Variable Stiffness Coil Assembly For Olympus CF-H190L Colonoscope ►Brief Description The Variable Stiffness Wire/Coil is a core mechanical element of Olympus’s proprietary RIT (Responsive Insertion Technology) found in the CF-H190L colonoscope. Unlike gastroscopes, colonoscopes navigate acute loops ...